User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Fluke 87:TRMS Multimeter DATE: 26-Sep-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 102 NUMBER OF LINES: 336 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # Source: Fluke 80 Series Service Manual (PN 834168 April 1989). # This procedure is run with the MET/CAL Configuration #3 loaded. 1.001 ASK- R N F W 1.002 ASK+ K X 1.003 HEAD * INITIAL CONDITIONS * 1.004 DISP The following Standard test equipment is needed for 1.004 DISP this procedure: 1.004 DISP GenRad Standard Capacitor 1409-Y 1.0µF 1.004 DISP GenRad Standard Capacitor 1409-X 0.5µF 1.004 DISP GenRad Standard Capacitor 1409-R 0.05µF 1.004 DISP GenRad Standard Capacitor 1409-K 0.005µF 1.005 DISP [32] [27][91]1m[27][91]5mWARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD {* DISPLAY TEST *} 1.007 JMP 2.001 1.008 EVAL 2.001 HEAD {Are all the display segments illuminated?} 2.002 PICE 87_dsp 3.001 HEAD 3.002 DISP No warm-up time requirement. 3.002 DISP Ambient temperature: 18C - 28C. 3.002 DISP Relative humidity: less than 90%. 3.003 HEAD {* ROTARY SWITCH TEST *} 3.004 JMP 4.001 3.005 EVAL 4.001 DISP Hold down the RANGE button while turning the UUT rotary 4.001 DISP function switch from OFF to VDC. Release the RANGE 4.001 DISP button and turn the switch to {VAC}. 4.002 EVAL {Does the UUT Indicate between -0012 and 0012?} 5.001 DISP Rotate the UUT function switch to {VDC}. 5.002 EVAL {Does the UUT Indicate between -0020 and -0044?} 6.001 DISP Rotate the UUT function switch to {mVDC}. 6.002 EVAL {Does the UUT Indicate between -0052 and -0076?} 7.001 DISP Rotate the UUT function switch to {Ohm}. 7.002 EVAL {Does the UUT Indicate between -0084 and -0108?} 8.001 DISP Rotate the UUT function switch to {Diode}. 8.002 EVAL {Does the UUT Indicate between -0116 and -0140?} 9.001 DISP Rotate the UUT function switch to {mA/A}. 9.002 EVAL {Does the UUT Indicate between -0148 and -0172?} 10.001 DISP Rotate the UUT function switch to {µA}. 10.002 EVAL {Does the UUT Indicate between -0180 and -0204?} 11.001 DISP Rotate the rotary switch to OFF. 11.002 HEAD {* AC VOLTAGE TEST *} 11.003 JMP 12.001 11.004 EVAL 12.001 DISP Rotate the UUT function switch to VAC. 12.002 DISP [32] Connect the 5700A and the UUT as follows. 12.002 DISP [32][27][91]1m 5700A TO Fluke 87 12.002 DISP [32] OUTPUT V Ohm A HI ......... V Ohm DIODE 12.002 DISP [32] OUTPUT V Ohm A LO ......... COM 12.003 HEAD AC VOLTAGE TESTS: {400mV Range} 12.004 JMP 13.001 12.005 EVAL 13.001 5700 350.0mV -2.8U +2.9U 60H 2W #! Test Tol 0.0028, Sys Tol 3.325e-005, TUR 84.211 (>= 4.00). 14.001 5700 350.0mV 3.9U 5kH 2W #! Test Tol 0.0039, Sys Tol 3.325e-005, TUR 117.293 (>= 4.00). 15.001 5700 350.0mV 7.4U 20kH 2W #! Test Tol 0.0074, Sys Tol 3.325e-005, TUR 222.556 (>= 4.00). 16.001 HEAD AC VOLTAGE TESTS: {4V Range} 16.002 JMP 17.001 16.003 EVAL 17.001 5700 3.500V -.026U +.027U 60H 2W #! Test Tol 0.026, Sys Tol 0.0003325, TUR 78.195 (>= 4.00). 18.001 5700 3.500V 0.039U 5kH 2W #! Test Tol 0.039, Sys Tol 0.0003325, TUR 117.293 (>= 4.00). 19.001 5700 3.500V 0.074U 20kH 2W #! Test Tol 0.074, Sys Tol 0.0003325, TUR 222.556 (>= 4.00). 20.001 HEAD AC VOLTAGE TESTS: {40V Range} 20.002 JMP 21.001 20.003 EVAL 21.001 5700 35.00V -0.26U +0.27U 60H 2W #! Test Tol 0.26, Sys Tol 0.0038, TUR 68.421 (>= 4.00). 22.001 5700 35.00V 0.39U 5kH 2W #! Test Tol 0.39, Sys Tol 0.0038, TUR 102.632 (>= 4.00). 23.001 5700 35.00V 0.74U 20kH 2W #! Test Tol 0.74, Sys Tol 0.0038, TUR 194.737 (>= 4.00). 24.001 HEAD AC VOLTAGE TESTS: {400V Range} 24.002 JMP 25.001 24.003 EVAL 25.001 5700 100.0V 2.4U 20kH 2W #! Test Tol 2.4, Sys Tol 0.009, TUR 266.667 (>= 4.00). 26.001 5700 200.0V 4.4U 20kH 2W #! Test Tol 4.4, Sys Tol 0.017, TUR 258.824 (>= 4.00). 27.001 5700 350.0V -2.6U +2.7U 60H B1 2W #! Test Tol 2.6, Sys Tol 0.032, TUR 81.250 (>= 4.00). 28.001 5700 350.0V 3.9U 5kH B1 2W #! Test Tol 3.9, Sys Tol 0.04975, TUR 78.392 (>= 4.00). 29.001 5700 300.0V 6.4U 20kH B1 2W #! Test Tol 6.4, Sys Tol 0.0435, TUR 147.126 (>= 4.00). 30.001 5700 350.0V 7.4U 20kH B1 2W #! Test Tol 7.4, Sys Tol 0.04975, TUR 148.744 (>= 4.00). 31.001 HEAD AC VOLTAGE TESTS: {1000V Range} 31.002 JMP 32.001 31.003 EVAL 32.001 5700 1000V 9U 60H B1 2W #! Test Tol 9, Sys Tol 0.084, TUR 107.143 (>= 4.00). 33.001 5700 1000V 14U 5kH B1 2W #! Test Tol 14, Sys Tol 0.131, TUR 106.870 (>= 4.00). 34.001 DISP Disconnect the 5700A from the UUT. 34.002 CALL Sub Fluke 83,85,87 Freq & Duty Cycle (1 year) CAL VER 34.003 JMP 35.001 34.004 EVAL dummy 35.001 HEAD {* DC VOLTAGE TEST *} 35.002 ASK+ K 35.003 JMP 36.001 35.004 EVAL 36.001 DISP Rotate the UUT function switch to VDC. 36.002 DISP [32] Connect the 5700A and the UUT as follows. 36.002 DISP [32][27][91]1m 5700A TO Fluke 87 36.002 DISP [32] OUTPUT V Ohm A HI ......... V Ohm DIODE 36.002 DISP [32] OUTPUT V Ohm A LO ......... COM 36.002 DISP [32] SENSE V Ohm HI ........... V Ohm DIODE 36.002 DISP [32] SENSE V Ohm HI ........... COM 36.003 HEAD DC VOLTAGE TESTS: {4V Range} 36.004 JMP 37.001 36.005 EVAL 37.001 5700 3.500V 0.005U 2W #! Test Tol 0.005, Sys Tol 2.15e-005, TUR 232.558 (>= 4.00). 38.001 HEAD DC VOLTAGE TESTS: {40V Range} 38.002 JMP 39.001 38.003 EVAL 39.001 5700 35.00V 0.05U 2W #! Test Tol 0.05, Sys Tol 0.00031, TUR 161.290 (>= 4.00). 40.001 5700 -35.00V 0.05U 2W #! Test Tol 0.05, Sys Tol 0.00031, TUR 161.290 (>= 4.00). 41.001 HEAD DC VOLTAGE TESTS: {400V Range} 41.002 JMP 42.001 41.003 EVAL 42.001 5700 350.0V 0.5U 2W #! Test Tol 0.5, Sys Tol 0.0034, TUR 147.059 (>= 4.00). 43.001 HEAD DC VOLTAGE TESTS: {1000V Range} 43.002 JMP 44.001 43.003 EVAL 44.001 5700 1000V 2U 2W #! Test Tol 2, Sys Tol 0.0086, TUR 232.558 (>= 4.00). 45.001 HEAD {* 1ms MIN/MAX TEST *} 45.002 JMP 46.001 45.003 EVAL 46.001 5700 2V 60H S 2W 46.002 DISP Press the MIN MAX button. 46.002 DISP Press the BEEPER button to display maximum values. 46.003 ACC 2.828VP 0.4% 2.05e-4U 46.004 MEMI Use keyboard to enter actual UUT reading in Volts: 46.005 MEME 46.006 MEMC V 0.045U #! WARNING: Test Tol 0.045, Sys Tol 0.011517, TUR 3.907 (< 4.00). 47.001 5700 2V 60H S 2W 47.002 DISP Press the MIN MAX button to display minimum values. 47.003 ACC -2.828VP 0.4% 2.05e-4U 47.004 MEMI Use keyboard to enter actual UUT reading in Volts: 47.005 MEME 47.006 MEMC V 0.045U #! WARNING: Test Tol 0.045, Sys Tol 0.011517, TUR 3.907 (< 4.00). 48.001 HEAD {* DC MILLIVOLT TEST *} 48.002 JMP 49.001 48.003 EVAL 49.001 DISP Rotate the UUT function switch to mVDC. 49.002 HEAD DC MILLIVOLT TEST: {400mV Range} 49.003 JMP 50.001 49.004 EVAL 50.001 5700 350.0mV 0.5U 2W #! Test Tol 0.0005, Sys Tol 3.3e-006, TUR 151.515 (>= 4.00). 51.001 HEAD {* RESISTANCE TEST *} 51.002 JMP 52.001 51.003 EVAL 52.001 DISP Rotate the UUT function switch to Ohm. 52.002 HEAD RESISTANCE TESTS: {400Ohm Range} 52.003 JMP 53.001 52.004 EVAL 53.001 5700 190.0Z 0.5U CW #! Test Tol 0.5, Sys Tol 0.00523, TUR 95.602 (>= 4.00). 54.001 HEAD RESISTANCE TESTS: {40kOhm Range} 54.002 JMP 55.001 54.003 EVAL 55.001 5700 19.00kZ 0.05U CW #! Test Tol 50, Sys Tol 0.309, TUR 161.812 (>= 4.00). 56.001 HEAD RESISTANCE TESTS: {4MOhm Range} 56.002 JMP 57.001 56.003 EVAL 57.001 5700 1.900MZ 0.005U 2W #! Test Tol 5000, Sys Tol 36.1, TUR 138.504 (>= 4.00). 58.001 HEAD RESISTANCE TESTS: {40MOhm Range} 58.002 JMP 59.001 58.003 EVAL 59.001 5700 19.00MZ 0.22U 2W #! Test Tol 2.2e+005, Sys Tol 893, TUR 246.361 (>= 4.00). 60.001 HEAD {* CONDUCTANCE TEST *} 60.002 JMP 61.001 60.003 EVAL 61.001 DISP Press the Range button twice for 40nS range. 61.002 HEAD CONDUCTANCE TEST: {40nS Range} 61.003 JMP 62.001 61.004 EVAL 62.001 5700 10nY 1.0% 0.10U 2W #! Test Tol 2e-010, Sys Tol 1.2001e-012, TUR 166.647 (>= 4.00). 63.001 HEAD {* CAPACITANCE TEST *} 63.002 JMP 64.001 63.003 EVAL 64.001 CALL Sub Fluke 83,85,87 Capacitance (1 year) CAL VER 64.002 HEAD {* DIODE TEST *} 64.003 JMP 65.001 64.004 EVAL 65.001 DISP [32] Connect the 5700A and the UUT as follows. 65.001 DISP [32][27][91]1m 5700A TO Fluke 87 65.001 DISP [32] OUTPUT V Ohm A HI ......... V Ohm DIODE 65.001 DISP [32] OUTPUT V Ohm A LO ......... COM 65.002 DISP Rotate the UUT function switch to Diode. 65.003 5700 3.000V 0.061U 2W #! Test Tol 0.061, Sys Tol 1.9e-005, TUR 3210.526 (>= 4.00). 66.001 HEAD {* DC MILLIAMP TEST *} 66.002 JMP 67.001 66.003 EVAL 67.001 DISP Rotate the UUT function switch to mA/A. 67.002 DISP [32] Connect the 5700A and the UUT as follows. 67.002 DISP [32][27][91]1m 5700A TO Fluke 87 67.002 DISP [32] OUTPUT V Ohm A HI ......... mA/µA 67.002 DISP [32] OUTPUT V Ohm A LO ......... COM 67.003 HEAD DC MILLIAMP TESTS: {40mA Range} 67.004 JMP 68.001 67.005 EVAL 68.001 5700 35.00mA 0.09U 2W #! Test Tol 9e-005, Sys Tol 3.1e-006, TUR 29.032 (>= 4.00). 69.001 HEAD DC MILLIAMP TESTS: {400mA Range} 69.002 JMP 70.001 69.003 EVAL 70.001 5700 350.0mA 0.9U 2W #! Test Tol 0.0009, Sys Tol 5.8e-005, TUR 15.517 (>= 4.00). 71.001 HEAD {* AC MILLIAMP TEST *} 71.002 JMP 72.001 71.003 EVAL 72.001 DISP Press the BLUE button to select AC. 72.002 HEAD AC MILLIAMP TESTS: {40mA Range} 72.003 JMP 73.001 72.004 EVAL 73.001 5700 35.00mA 0.37U 60H 2W #! Test Tol 0.00037, Sys Tol 9.25e-006, TUR 40.000 (>= 4.00). 74.001 5700 35.00mA 0.37U 1kH 2W #! Test Tol 0.00037, Sys Tol 9.25e-006, TUR 40.000 (>= 4.00). 75.001 HEAD AC MILLIAMP TESTS: {400mA Range} 75.002 JMP 76.001 75.003 EVAL 76.001 5700 350.0mA 3.7U 60H 2W #! Test Tol 0.0037, Sys Tol 0.0002675, TUR 13.832 (>= 4.00). 77.001 5700 350.0mA 3.7U 1kH 2W #! Test Tol 0.0037, Sys Tol 0.0002675, TUR 13.832 (>= 4.00). 78.001 HEAD {* DC MICROAMP TEST *} 78.002 JMP 79.001 78.003 EVAL 79.001 DISP Rotate the UUT function switch to µA. 79.002 HEAD DC MICROAMP TESTS: {400µA Range} 79.003 JMP 80.001 79.004 EVAL 80.001 5700 350.0uA 0.9U 2W #! Test Tol 9e-007, Sys Tol 2.75e-008, TUR 32.727 (>= 4.00). 81.001 HEAD DC MICROAMP TESTS: {4000µA Range} 81.002 JMP 82.001 81.003 EVAL 82.001 5700 3500uA 9U 2W #! Test Tol 9e-006, Sys Tol 2.75e-007, TUR 32.727 (>= 4.00). 83.001 HEAD {* AC MICROAMP TEST *} 83.002 JMP 84.001 83.003 EVAL 84.001 DISP Press the BLUE button to select AC. 84.002 HEAD AC MICROAMP TESTS: {400µA Range} 84.003 JMP 85.001 84.004 EVAL 85.001 5700 350.0uA 3.7U 60H 2W #! Test Tol 3.7e-006, Sys Tol 8.9e-008, TUR 41.573 (>= 4.00). 86.001 5700 350.0uA 3.7U 1kH 2W #! Test Tol 3.7e-006, Sys Tol 8.9e-008, TUR 41.573 (>= 4.00). 87.001 HEAD AC MICROAMP TESTS: {4000µA Range} 87.002 JMP 88.001 87.003 EVAL 88.001 5700 3500uA 37U 60H 2W #! Test Tol 3.7e-005, Sys Tol 8.9e-007, TUR 41.573 (>= 4.00). 89.001 5700 3500uA 37U 1kH 2W #! Test Tol 3.7e-005, Sys Tol 8.9e-007, TUR 41.573 (>= 4.00). 90.001 DISP Disconnect the 5700A from the UUT. 90.002 HEAD {* DC AMP TEST *} 90.003 JMP 91.001 90.004 EVAL 91.001 DISP Rotate the UUT function switch to mA/A. 91.002 DISP [32]Connect the 5725A and UUT as follows: 91.002 DISP [32][27][91]1m 5725A TO Fluke 87 91.002 DISP [32] CURRENT OUTPUT HI ............ A 91.002 DISP [32] CURRENT OUTPUT LO ............ COM 91.003 HEAD DC AMP TESTS: {4000mA Range} 91.004 JMP 92.001 91.005 EVAL 92.001 5700 3500mA 9U B1 2W #! Test Tol 0.009, Sys Tol 0.00167, TUR 5.389 (>= 4.00). 93.001 HEAD DC AMP TESTS: {10A Range} 93.002 JMP 94.001 93.003 EVAL 94.001 5700 10.00A 0.04U B1 2W #! Test Tol 0.04, Sys Tol 0.00388, TUR 10.309 (>= 4.00). 95.001 HEAD {* AC AMP TEST *} 95.002 JMP 96.001 95.003 EVAL 96.001 DISP Press the BLUE button to select AC. 96.002 HEAD AC AMP TESTS: {4000mA Range} 96.003 JMP 97.001 96.004 EVAL 97.001 5700 3500mA 37U 60H B1 2W #! Test Tol 0.037, Sys Tol 0.00157, TUR 23.567 (>= 4.00). 98.001 5700 3500mA 37U 1kH B1 2W #! Test Tol 0.037, Sys Tol 0.00157, TUR 23.567 (>= 4.00). 99.001 HEAD AC AMP TESTS: {10A Range} 99.002 JMP 100.001 99.003 EVAL 100.001 5700 10.00A 0.12U 60H B1 2W #! Test Tol 0.12, Sys Tol 0.00417, TUR 28.777 (>= 4.00). 101.001 5700 10.00A 0.12U 1kH B1 2W #! Test Tol 0.12, Sys Tol 0.00417, TUR 28.777 (>= 4.00). 102.001 END #! T.U.R.s less than 4.00: 2 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.